Frequently Asked Question

Analyzer out of Range
Zuletzt aktualisiert vor 5 Jahren

This message can happen during CO-Diffusion measurement.

Please check:

- if power is switched on

- if 10 Minutes Warmup Time is past

When this is ok start with the next settings:

Open smartSOFTmee, Setup->Configuration->Device->DLCO

image

Change MinOffsetCO and CH4 to "0", press safe and restart the Software