Frequently Asked Question

TGV Shutter close Time/breath
Zuletzt aktualisiert vor 5 Jahren

In our C:/MEE/SMARTSOFTMEE/device.cfg file you can make the setting if the Shutter is closed for a fix time or for a number of breathes.

[BODY]
TGVShutterCloseTime=3.0 ; s
TGVShutterCloseBreaths=3 ; 0: Fixed shutter close time, else: number of breaths